Turbidity and Chlorine (Free & Total) Handheld Analyzer - One Meter, Three Measurements

Hanna HI93414: Measuring the most important parameters in drinking water.

Three options for measurement including ratio mode for turbidity, free chlorine, and total chlorine.

Turbidity and Chlorine (Free & Total) Portable Metre.

The Hanna Turbidity and Chlorine HI93414 is a high accuracy Portable Meter. Combining nephelometer and photometer to measure key parameters in drinking water.

calibration and performance verification is ensured with the supplied AMCO-AEPA-1 primary turbidity standards and NIST traceable secondary chlorine standards.

The HI93414 meets and exceeds the requirements of EPA Method 180.1 and Standard Methods for the Examination of Water and Wastewater 2130 B for turbidity measurements.The chlorine measurements are are adapted EPA Method 330.5 and Standard Method 4500-Cl G.

Hanna Instrument Features:

  • USB for data transfer.
  • Operates on batteries or line voltage.
  • Three Measurement Modes – The HI93414 features three options for measurement including ratio mode for turbidity, free chlorine, and total chlorine. Turbidity measurements can be made in the 0.00 to 1000 NTU (Nephelometric Turbidity Units) range, while free or total chlorine measurements can be made in the 0.00 to 5.00 mg/L (ppm) range.
  • Multiple reading modes – Normal measurement, continuous measurement, or signal averaging measurement are reading modes available
  • EPA Compliant – The HI93414 meets and exceeds the requirements of EPA and Standard Methods both for turbidity and colorimetric chlorine measurements. When the meter is in EPA mode all turbidity readings are rounded accordingly to meet reporting requirements.
  • Calibration – A two, three, or four-point turbidity calibration can be performed by using the supplied (<0.10, 15.0, 100, and 750 NTU) standards. Calibration points can be modified if user-prepared standards are used. For free and total chlorine, the 1.00 mg/L (ppm) CAL Check™ standard can be used for calibration and performance verification.
  • AMCO AEPA-1 Primary Turbidity Standard – The AMCO AEPA-1 supplied standards are recognized as a primary standard by the USEPA. These non-toxic standards are made of styrene divinylbenzene polymer spheres that are uniform in size and density. The standards are reusable and stable with a long shelf life.
  • CAL Check™ – With the powerful CAL Check™ function, reliable performance of the chlorine colorimeter can be validated at any moment by using the exclusive HANNA ready-made, NIST traceable standards. All standards are supplied with a Certificate of Analysis (COA) for traceability.
  • Fast Tracker™ – For advanced ˜field applications, the HI93414 is equipped with Fast Tracker™ – Tag Identi˜fication System (T.I.S.) that makes data collecting and management simpler than ever. Fast Tracker™ allows users to record the time and location of a speci˜fic measurement or series of measurements using iButton® tags near sampling points for quick and easy readings. Each iButton® tag contains a computer chip with a unique identi˜cation code encased in stainless steel.
  • GLP Data – The HI93414 features complete GLP (Good Laboratory Practice) functions that allow traceability of the calibration conditions. Data includes calibration points, date, and time.
  • Data Logging – Up to 200 measurements can be stored in the internal memory and recalled at any time.
  • Data Transfer – For further storage or analysis options, logged data can be downloaded to a Windows compatible PC using the RS232 or USB ports and the HI92000 software.
  • Backlit Display – A backlit LCD display provides an easy to understand, user-friendly interface. Displayed codes guide the user step-by-step through routine operation and calibration.

Hanna Turbidity Calibration Standards

The HI98703-11 AMCO AEPA-1 standards ensure that measurements are traceable to a primary reference materials. These standards are used for calibration and performance verification of the HI93414 turbidity meter.

Supplied with Certificate of Analysis
  • Lot number.
  • Expiration date.
  • Standard value @ 25 °C.
  • Reference meter NIST traceable.
  • Provided storage containers.
  • Light tight.
  • Protects from accidental breakage.

Hanna Chlorine Calibration Standards

The HI93414-11 free and total chlorine CAL Check™standards are used for calibration and performance verification of the HI93414.

Supplied with Certificate of Analysis
  • Lot number.
  • Expiration date.
  • Standard value @ 25 °C.
  • Reference meter NIST traceable.
  • Provided storage containers.
  • Light tight.
  • Protects from accidental breakage.
Hanna Chlorine Turbidity Carry Case
Carry case for Instrument, Cuvettes & Standards etc.
Hanna Instrument Rear Closeup
Rear showing RS232, USB and ID Tag Reader
Hanna Chlorine Turbidity Front View
Hanna Chlorine & Turbidity Instrument Front
Hanna Instrument CAL Check™ standards with certificate
Hanna CAL Check™ Standards with Certificate
SKUHI93414-01
Product NameTurbidity (EPA) and Chlorine Portable Meter – HI93414
Quote RequiredNo
Turbidity Range0.00 to 9.99; 10.0 to 99.9; 100 to 1000 NTU
Turbidity Resolution0.01; 0.1; 1 NTU
Turbidity Accuracy±2% of reading plus 0.02 NTU
Turbidity Repeatability±1% of reading or 0.02 NTU, whichever is greater
Turbidity Stray Light< 0.02 NTU
Turbidity Calibrationtwo, three, or four-point calibration
Turbidity Measuring Methodratio nephelometric method (90° & 180°)
Turbidity Measuring Modesnormal, average, continuous
Turbidity Light Sourcetungsten filament lamp
Turbidity Light Detectorsilicon photocell
Logging Memory200 records
ConnectivityUSB or RS232
Automatic Shut-Offauto-off after 15 minutes of non-use
Display40 x 70 mm graphic LCD (64 x 128 pixels) with backlight
Battery Type/Life4 (1.5V) AA alkaline batteries
Power Supply4 (1.5V) AA alkaline batteries or AC adapter
Environment0 to 50°C (32 to 122°F), RH max 95% non-condensing
Dimensions224 x 87 x 77 mm (8.8 x 3.4 x 3.0”)
Weight512 g (18 oz.)
PhotoMeter/ColoriMeter Light SourceTungsten filament lamp
PhotoMetre/ColoriMeter Light DetectorSilicon photocell with 525 nm narrow band interference filters
Ordering InformationHI93414 is supplied with sample cuvettes and caps (5), calibration cuvettes for turbidity and chlorine, silicone oil, cuvette wiping cloth, scissors, batteries, AC adapter, instruction manual, and rugged carrying case.
Free Chlorine Range0.00 to 5.00 mg/L
Free Chlorine Resolution0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 above 3.50 mg/L
Free Chlorine Accuracy±0.02 mg/L @ 1.00 mg/L
Total Chlorine Range0.00 to 5.00 mg/L
Total Chlorine Resolution0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 above 3.50 mg/L
Total Chlorine Accuracy±0.02 mg/L @ 1.00 mg/L
Chlorine MethodAdaptation of the USEPA Method 330.5 and Standard Method 4500-Cl G. The reaction between chlorine and DPD reagent causes a pink tint in the sample.
Description

The Hanna Turbidity and Chlorine HI93414 is a high accuracy Portable Meter. Combining nephelometer and photometer to measure key parameters in drinking water.

calibration and performance verification is ensured with the supplied AMCO-AEPA-1 primary turbidity standards and NIST traceable secondary chlorine standards.

The HI93414 meets and exceeds the requirements of EPA Method 180.1 and Standard Methods for the Examination of Water and Wastewater 2130 B for turbidity measurements.The chlorine measurements are are adapted EPA Method 330.5 and Standard Method 4500-Cl G.

Hanna Instrument Features:

  • USB for data transfer.
  • Operates on batteries or line voltage.
  • Three Measurement Modes – The HI93414 features three options for measurement including ratio mode for turbidity, free chlorine, and total chlorine. Turbidity measurements can be made in the 0.00 to 1000 NTU (Nephelometric Turbidity Units) range, while free or total chlorine measurements can be made in the 0.00 to 5.00 mg/L (ppm) range.
  • Multiple reading modes – Normal measurement, continuous measurement, or signal averaging measurement are reading modes available
  • EPA Compliant – The HI93414 meets and exceeds the requirements of EPA and Standard Methods both for turbidity and colorimetric chlorine measurements. When the meter is in EPA mode all turbidity readings are rounded accordingly to meet reporting requirements.
  • Calibration – A two, three, or four-point turbidity calibration can be performed by using the supplied (<0.10, 15.0, 100, and 750 NTU) standards. Calibration points can be modified if user-prepared standards are used. For free and total chlorine, the 1.00 mg/L (ppm) CAL Check™ standard can be used for calibration and performance verification.
  • AMCO AEPA-1 Primary Turbidity Standard – The AMCO AEPA-1 supplied standards are recognized as a primary standard by the USEPA. These non-toxic standards are made of styrene divinylbenzene polymer spheres that are uniform in size and density. The standards are reusable and stable with a long shelf life.
  • CAL Check™ – With the powerful CAL Check™ function, reliable performance of the chlorine colorimeter can be validated at any moment by using the exclusive HANNA ready-made, NIST traceable standards. All standards are supplied with a Certificate of Analysis (COA) for traceability.
  • Fast Tracker™ – For advanced ˜field applications, the HI93414 is equipped with Fast Tracker™ – Tag Identi˜fication System (T.I.S.) that makes data collecting and management simpler than ever. Fast Tracker™ allows users to record the time and location of a speci˜fic measurement or series of measurements using iButton® tags near sampling points for quick and easy readings. Each iButton® tag contains a computer chip with a unique identi˜cation code encased in stainless steel.
  • GLP Data – The HI93414 features complete GLP (Good Laboratory Practice) functions that allow traceability of the calibration conditions. Data includes calibration points, date, and time.
  • Data Logging – Up to 200 measurements can be stored in the internal memory and recalled at any time.
  • Data Transfer – For further storage or analysis options, logged data can be downloaded to a Windows compatible PC using the RS232 or USB ports and the HI92000 software.
  • Backlit Display – A backlit LCD display provides an easy to understand, user-friendly interface. Displayed codes guide the user step-by-step through routine operation and calibration.

Hanna Turbidity Calibration Standards

The HI98703-11 AMCO AEPA-1 standards ensure that measurements are traceable to a primary reference materials. These standards are used for calibration and performance verification of the HI93414 turbidity meter.

Supplied with Certificate of Analysis
  • Lot number.
  • Expiration date.
  • Standard value @ 25 °C.
  • Reference meter NIST traceable.
  • Provided storage containers.
  • Light tight.
  • Protects from accidental breakage.

Hanna Chlorine Calibration Standards

The HI93414-11 free and total chlorine CAL Check™standards are used for calibration and performance verification of the HI93414.

Supplied with Certificate of Analysis
  • Lot number.
  • Expiration date.
  • Standard value @ 25 °C.
  • Reference meter NIST traceable.
  • Provided storage containers.
  • Light tight.
  • Protects from accidental breakage.
Hanna Chlorine Turbidity Carry Case
Carry case for Instrument, Cuvettes & Standards etc.
Hanna Instrument Rear Closeup
Rear showing RS232, USB and ID Tag Reader
Hanna Chlorine Turbidity Front View
Hanna Chlorine & Turbidity Instrument Front
Hanna Instrument CAL Check™ standards with certificate
Hanna CAL Check™ Standards with Certificate
Specifications
SKUHI93414-01
Product NameTurbidity (EPA) and Chlorine Portable Meter – HI93414
Quote RequiredNo
Turbidity Range0.00 to 9.99; 10.0 to 99.9; 100 to 1000 NTU
Turbidity Resolution0.01; 0.1; 1 NTU
Turbidity Accuracy±2% of reading plus 0.02 NTU
Turbidity Repeatability±1% of reading or 0.02 NTU, whichever is greater
Turbidity Stray Light< 0.02 NTU
Turbidity Calibrationtwo, three, or four-point calibration
Turbidity Measuring Methodratio nephelometric method (90° & 180°)
Turbidity Measuring Modesnormal, average, continuous
Turbidity Light Sourcetungsten filament lamp
Turbidity Light Detectorsilicon photocell
Logging Memory200 records
ConnectivityUSB or RS232
Automatic Shut-Offauto-off after 15 minutes of non-use
Display40 x 70 mm graphic LCD (64 x 128 pixels) with backlight
Battery Type/Life4 (1.5V) AA alkaline batteries
Power Supply4 (1.5V) AA alkaline batteries or AC adapter
Environment0 to 50°C (32 to 122°F), RH max 95% non-condensing
Dimensions224 x 87 x 77 mm (8.8 x 3.4 x 3.0”)
Weight512 g (18 oz.)
PhotoMeter/ColoriMeter Light SourceTungsten filament lamp
PhotoMetre/ColoriMeter Light DetectorSilicon photocell with 525 nm narrow band interference filters
Ordering InformationHI93414 is supplied with sample cuvettes and caps (5), calibration cuvettes for turbidity and chlorine, silicone oil, cuvette wiping cloth, scissors, batteries, AC adapter, instruction manual, and rugged carrying case.
Free Chlorine Range0.00 to 5.00 mg/L
Free Chlorine Resolution0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 above 3.50 mg/L
Free Chlorine Accuracy±0.02 mg/L @ 1.00 mg/L
Total Chlorine Range0.00 to 5.00 mg/L
Total Chlorine Resolution0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 above 3.50 mg/L
Total Chlorine Accuracy±0.02 mg/L @ 1.00 mg/L
Chlorine MethodAdaptation of the USEPA Method 330.5 and Standard Method 4500-Cl G. The reaction between chlorine and DPD reagent causes a pink tint in the sample.