TYPE B
SENSING ELEMENT: Ultra High-Capacitance Aluminium Oxide Type.
RANGE IN DEWPOINT:
- SR: -166°F to -4°F (-110°C to -20°C) dewpoint
- PL: -148°F to +32°F (-100°C to 0°C) dewpoint
- RD: -112°F to -4°F (-80°C to -20°C) dewpoint
- GY: -112°F to +32°F (-80°C to 0°C) dewpoint
- BL: -112°F to +68°F (-80°C to +20°C) dewpoint
DISPLAY UNITS:
- °F – Degrees Fahrenheit dew / frost Point
- °C – Degrees Centigrade dew / frost Point
- ppm(v) – Parts per million (volume)
- ppb(v) – Parts per billion (volume)
- ppm(w) – Parts per million (weight)
- g/m3 – Grams per cubic meter
- lbs/MMSCF – Pounds per million standard cubic feet
DISPLAY: Blue on Green, 128 x 64 pixel, Graphical LCD with LED backlight.
SENSOR CALIBRATION ACCURACY: Better than ±3.6°F (±2°C) dewpoint. Each unit supplied with a Certificate of Calibration, traceable to National & International Standards – National Physical Laboratory (UK) / NIST (USA).
AUTOMATIC CALIBRATION: Electronic “Span Check”. Performed by user following simple menu driven instructions. Can be password protected to avoid unauthorized tamper.
REPEATABILITY: Better than ±0.36°F (±0.2°C) dewpoint.
POWER SUPPLY: Rechargeable Li-Ion Battery. Battery charger included.
BATTERY LIFE: In excess of 80 hours of continuous use on full charge at 68°F/20°C.
TIME TO FULL CHARGE: 14 hours.
KEYBOARD: 5 Membrane covered, metallic dome tactile keys.
PRESSURE CORRECTION: Integral calculator to display pressure dew points.
Gauge pressure can be entered in psi, bar, kPa or kg/cm2.
TEMPERATURE COEFFICIENT: Temperature compensated for operating range.
GAS SAMPLE CONNECTIONS: Ports accept Swagelok® VCO type coupling (9/16″ x 18 UNF). Supplied with either 1/4″ or 6mm Swagelok® SS compression fitting on one side. The other side is fitted with a stainless steel push-on, “fir tree” type, hose connector for 6mm ID tube. Each unit is supplied with a 6.6ft (2m) length of 6mm ID PTFE tube.
ELECTRICAL CONNECTIONS: 9 Pin “D” type for 4-20mA analogue output, RS485 Serial Communications and PC interface. Separate socket for battery charger.
OPERATING TEMPERATURE: -4°F to +122°F (-20°C to +50°C).
STORAGE TEMPERATURE & HUMIDITY: -40°F to +176°F (-40°C to +80°C) / 95% RH Non-condensing.
OPERATING PRESSURE: Atmospheric pressure, maximum 0.3barg / 4 psig (30kPag).
OPERATING HUMIDITY (External): 95% RH Non-condensing.
TYPICAL RESPONSE TIMES:
- Wet to Dry: -14°F to -76°F (-10°C to -60°C) – less than 120 seconds.
- Dry to Wet: -166°F to -4°F (-110°C to -20°C) – less than 20 seconds
SAMPLE FLOW RATE: Flow independent, but ideally 2 to 5 litres per minute. Max: 10 litres/min.
REPLACEMENT DESICCANT: Field Interchangeable.
SENSOR LIFE: Between 5 & 10 years – depending on application.
REPLACEMENT SENSOR: Field Interchangeable.
ELECTROMAGNETIC COMPATIBILITY (EMC): Product complies with the objectives and requirements of EMC Directive 2014/30/EU, 2014/34/EU, EN 60079-0:2009, EN 60079-11:2012, EN 60079-26:2007 & BS EN 61326-1:2006.
SECURITY: Multi level password protection.
WARM UP TIME: 10 seconds.
WEATHERPROOF CLASSIFICATION: IP54 / NEMA12.
WARRANTY: 2 years for faulty workmanship and defective parts.
WEIGHT: 2.65lbs (1.2kgs).
DIMENSIONS: 7.5 x 3.8 x 4.5 inches (192 x 97 x 115 mms).
MATERIALS OF CONSTRUCTION: Sensor in metal housing. Outer case custom manufactured in, stainless steel impregnated, high impact Polybutylene Terephthalate (PBT).
ACCESSORIES INCLUDED: Anti-static carrying case for use in hazardous areas. 2m PTFE Sampling pipe, Universal Battery Charger, User Manual.
TYPE L OPTION
As TYPE B ABOVE plus following additional specifications and
features:
DATA LOGGING: 16,000 samples. Date and time stamped data, stored in chosen units of measurement for download to PC.
DATA LOCATION: 20 separate locations (Tags) can be entered (alpha-numerical) by user for data collection at pre-programmed locations.
SAMPLING RATES: User selectable (in intervals of 6 seconds) from once every 6 seconds to once a day.
DATA DISPLAY: Numerical and Graphical display of data on SADPmini2 -Ex screen.
REAL TIME RECORDING: Device can be programmed to monitor, record and graphically present data in real time directly to PC.
SERIAL COMMUNICATIONS: RS485, baud rate 9600 – half duplex.
ANALOGUE OUTPUT: Externally powered 4-20mA loop. Linear output with unit selected. Span easily configured by user.
Max. load = 50 x (VEXT – 0.6) – 105
VEXT = Supply voltage.
(Eg. For 24V supply, Max. load = 1065Ω)
ACCESSORIES INCLUDED: 4-20mA / RS485 Connector for analogue output and serial/USB communications. Serial or USB (Isolated / self powered) interface, with cables, for “Real Time Logging” and data download to PC. Software supplied on CD.
SYSTEM REQUIREMENT: Windows® XP or later for Serial USB connection to
a PC