wjf-logo
1 877 291 5572    |    Contact Us    |    Search

Online Phosphate Analyzer

Online Phosphate Analyzer using the colorimetric process for the automatic, continuous measurement of phosphate.

Online Phosphate Analyzer
  • Description
  • Specification
  • Application
  • Options
  • Downloads
  • F.A.Q

The Online Phosphate Analyzer from SWAN Analytical Instruments’. The AMI Phosphate-II Measuring range: 0.01 to 10 ppm (mg/L) PO4.

Monitor AMI Phosphate-II

  • Measuring range: 0.01 to 10 ppm (mg/L) PO4
  • Based on colorimetric measurement principle according to EN ISO 6878 / APHA 4500 –P E.
  • No interferences with silica.
  • Complete system including measurement and control electronics, photometer, flow indicator, reaction chamber, reagent dosing system and reagent containers.
  • Measurement values are available as analog output signals.
  • Alarm display and activation of alarm relay when user defined, critical limits are reached.
  • Continuous, automatic monitoring of main instrument functions (sample flow, reagent supply).
  • Large back-lit LC display showing all measured values and status information simultaneously.
  • Easy user menus in English, German, French and Spanish. Simple programming of all parameters by keypad.
  • Data logger for 1’500 data records stored at a selectable interval. Factory tested, ready for installation and operation.
Online Phosphate Analyzer

Analyzer Specifications

Phosphate (PO4) Measurement
Measuring range:
– Standard operating mode
– Extended operating mode
– 0.1 to 25 ppm
– 0.1 to 50 ppm
Reproducibility± 0.01 ppm or ± 2.5%
whichever is greater
Measurement time2.5 minutes
Cycle time5 minutes up to 30 minutes
Power Supply
Voltage100-240 VAC (±10%), 50/60Hz
or
10-36 VDC
Sample conditions
Flow rateapprox. 10 l/h
Temperatureup to 50° C
Inlet pressure0.15 to 2 bar
Outlet pressurepressure free
Dimensions
Panel400 x 850 x 200 mm

For further information see the full Online Phosphate Analyzer Data Sheet.

Potable Water
For monitoring of permissible limits of phosphate in drinking water.

Effluents
High levels can lead to eutrophication, the depletion of oxygen from excessive aquatic growth.

Cooling Water
Manage accurate Phosphate levels to protect equipment from scale buildup.

Consumables

  • A-85.143.400 Phosphate standard solution 1000ppm, bottle with 100ml
  • A-85.420.660 Reagent set Oxycon On-line Phosphate
  • A-86.191.151 Tube 2.29mm in frame for PeriClip pump

 

Online Phosphate Analyzer Accessories

  • A-81.420.020 AMI Option; Profibus- / Modbus Interface
  • A-81.420.042 AMI Option; USB Interface
  • A-81.420.050 AMI Option; 3rd signal output
  • A-82.312.000 Cleaning module-II
  • A-85.151.100 VERIFICATION-KIT AMI Photometer, short
  • A-86.180.010 Sample tubing, LD-PE, 6×4 mm, 3m
  • A-88.224.220 Communication cable, AMI to PC, 2 m
  • A-89.812.100 Relay box for AMI transmitters

 

Online Phosphate Analyzer Spare Parts

  • A-82.931.060 Reagent container 2.5l with cover
  • A-83.590.030 Cleaning set for Trides flow cell
  • A-83.590.036 Spare part set for Codes-II flow cell and fotometer
  • A-86.191.871 Tubing kit to suction lance; V2
    A-86.191.890 Tubing kit reagent AMI Codes-II and AMI Phosphate-II
  • A-89.811.060 Set of fuses for AMI
  • Transmitter temperature monitoring with programmable high/low alarm limits
  • 1 Alarm relay One potential free contact for summary alarm indication for programmable alarm values and instrument faults. Maximum load: 1A / 250 VAC
  • 1 Input One input for potential-free contact. Programmable hold or remote off function.
  • 2 Relay outputs Two potential-free contacts programmable as limit switches for measuring values, controllers or timer for system cleaning with automatic hold function. Rated load: 1A / 250 VAC
  • 2 Signal outputs (3rd as option)Two programmable signal outputs for measured values (freely scalable, linear or bilinear) or as continuous control outputs (control parameters programmable) as current source.
  • 3rd signal output selectable as cur-rent source or current sink. Current loop: 0/4 – 20 mA Maximum burden: 510 Control functions Relays or current outputs programmable for 1 or 2 pulse dosing pumps, solenoid valves or for one motor valve. Programmable P, PI, PID or PD control parameters.
  • 1 Communication interface (option) RS485 interface (galvanically separated) with Fieldbus protocol Modbus RTU or Pro-fibus DP / 3rd Signal output / USB int

All Swan Analyzers are CSA Cirtified at our premises before shipping.

The Online Phosphate Analyzer from SWAN Analytical Instruments’. The AMI Phosphate-II Measuring range: 0.01 to 10 ppm (mg/L) PO4.

Monitor AMI Phosphate-II

  • Measuring range: 0.01 to 10 ppm (mg/L) PO4
  • Based on colorimetric measurement principle according to EN ISO 6878 / APHA 4500 –P E.
  • No interferences with silica.
  • Complete system including measurement and control electronics, photometer, flow indicator, reaction chamber, reagent dosing system and reagent containers.
  • Measurement values are available as analog output signals.
  • Alarm display and activation of alarm relay when user defined, critical limits are reached.
  • Continuous, automatic monitoring of main instrument functions (sample flow, reagent supply).
  • Large back-lit LC display showing all measured values and status information simultaneously.
  • Easy user menus in English, German, French and Spanish. Simple programming of all parameters by keypad.
  • Data logger for 1’500 data records stored at a selectable interval. Factory tested, ready for installation and operation.
Online Phosphate Analyzer

Analyzer Specifications

Phosphate (PO4) Measurement
Measuring range:
– Standard operating mode
– Extended operating mode
– 0.1 to 25 ppm
– 0.1 to 50 ppm
Reproducibility± 0.01 ppm or ± 2.5%
whichever is greater
Measurement time2.5 minutes
Cycle time5 minutes up to 30 minutes
Power Supply
Voltage100-240 VAC (±10%), 50/60Hz
or
10-36 VDC
Sample conditions
Flow rateapprox. 10 l/h
Temperatureup to 50° C
Inlet pressure0.15 to 2 bar
Outlet pressurepressure free
Dimensions
Panel400 x 850 x 200 mm

For further information see the full Online Phosphate Analyzer Data Sheet.

Potable Water
For monitoring of permissible limits of phosphate in drinking water.

Effluents
High levels can lead to eutrophication, the depletion of oxygen from excessive aquatic growth.

Cooling Water
Manage accurate Phosphate levels to protect equipment from scale buildup.

Consumables

  • A-85.143.400 Phosphate standard solution 1000ppm, bottle with 100ml
  • A-85.420.660 Reagent set Oxycon On-line Phosphate
  • A-86.191.151 Tube 2.29mm in frame for PeriClip pump

 

Online Phosphate Analyzer Accessories

  • A-81.420.020 AMI Option; Profibus- / Modbus Interface
  • A-81.420.042 AMI Option; USB Interface
  • A-81.420.050 AMI Option; 3rd signal output
  • A-82.312.000 Cleaning module-II
  • A-85.151.100 VERIFICATION-KIT AMI Photometer, short
  • A-86.180.010 Sample tubing, LD-PE, 6×4 mm, 3m
  • A-88.224.220 Communication cable, AMI to PC, 2 m
  • A-89.812.100 Relay box for AMI transmitters

 

Online Phosphate Analyzer Spare Parts

  • A-82.931.060 Reagent container 2.5l with cover
  • A-83.590.030 Cleaning set for Trides flow cell
  • A-83.590.036 Spare part set for Codes-II flow cell and fotometer
  • A-86.191.871 Tubing kit to suction lance; V2
    A-86.191.890 Tubing kit reagent AMI Codes-II and AMI Phosphate-II
  • A-89.811.060 Set of fuses for AMI
  • Transmitter temperature monitoring with programmable high/low alarm limits
  • 1 Alarm relay One potential free contact for summary alarm indication for programmable alarm values and instrument faults. Maximum load: 1A / 250 VAC
  • 1 Input One input for potential-free contact. Programmable hold or remote off function.
  • 2 Relay outputs Two potential-free contacts programmable as limit switches for measuring values, controllers or timer for system cleaning with automatic hold function. Rated load: 1A / 250 VAC
  • 2 Signal outputs (3rd as option)Two programmable signal outputs for measured values (freely scalable, linear or bilinear) or as continuous control outputs (control parameters programmable) as current source.
  • 3rd signal output selectable as cur-rent source or current sink. Current loop: 0/4 – 20 mA Maximum burden: 510 Control functions Relays or current outputs programmable for 1 or 2 pulse dosing pumps, solenoid valves or for one motor valve. Programmable P, PI, PID or PD control parameters.
  • 1 Communication interface (option) RS485 interface (galvanically separated) with Fieldbus protocol Modbus RTU or Pro-fibus DP / 3rd Signal output / USB int

All Swan Analyzers are CSA Cirtified at our premises before shipping.