wjf-logo
1 877 291 5572    |    Contact Us    |    Search

Iron Analyzer (Online) - Total Dissolved Iron

Seres Topaz Iron Analyzer

Continuous analysis of dissolved iron Fe (II+III) in potable water, surface water, wastewater and effluents.

Seres OL Analyzer Topaz Iron
  • Description
  • Specification
  • Application
  • Options
  • Downloads
  • F.A.Q

Total Dissolved Iron Anlysis

Complete iron analyzer system for the automatic, continuous analysis of dissolved iron Fe (II+III) in potable water, surface water, wastewater and effluents.

For the continuous, colorimetric online determination of dissolved iron Fe (II+III) per ISO 6332:1988, DIN 38406-E1-1 and NF T90 017.

  • Available in separate measuring ranges:
    • Seres OL Topaz Iron LR:
      • 0 – 100 ppb or
      • 0 – 1000 ppb
    • Seres OL Topaz Iron HR:
      • 0 – 2 ppm or
      • 0 – 5 ppm
  • Complete iron analyzer including measurement and control electronics, measuring unit, flow indicator, reaction chamber and reagent dosing system.
  • Robust, high quality analyzer cabinet painted stainless steel, 316L.
  • Automatic, electrical zero measurement prior to each measurement cycle.
  • Automatic cell cleaning.
  • 3 easily accessible peristaltic pump modules for accurate, automatic dosing of chemical reagents.
  • 2 analog and 7 digital outputs for alarms for process values and diagnostic alarms for each sample stream.
  • RS485 Modbus/JBUS RTU interface.
  • Large back-lit touchscreen color LCD display for the reading of all measured values and status information simultaneously.
  • Easy menu-guided operation in English or French.

The Serez OL Topaz series provides you with a high-quality industrial-grade analyzer specific to your Dissolved Iron analysis requirement.

Fluoride

Colorimetric method; in a medium buffered to pH 4.6, iron III is reduced to iron II. Phenanthroline then forms a complex with iron II of red color which is detected by the iron analyzer.

  • Reaction time 15-18 min.
  • Topaz Iron LR 0-100 or 0-1000 ppb
    • Limit of Detection 8 ppb
    • Repeatability < ± 2 % FS or ± 0.008ppm (whichever is greater)
    • Precision < ± 2 % FS or ± 0.008ppm (whichever is greater)
  • Topaz Iron HR 0-2 or 0-5 ppm
    • Limit of Detection 8 ppb
    • Repeatability < ± 2 % FS
    • Precision < ± 2 % FS


Complete iron analyzer including measurement and control electronics, measuring unit, flow indicator, reaction chamber and reagent dosing system.

Signal outputs

  • 2 programmable signal outputs for measured values (freely scalable, linear).
  • Current loop: 4 – 20 mA
  • Communication interfaceRS485 interface (galvanically separated) with Modbus/JBUS RTU protocol.
  • Ethernet interface (TCP/IP) optional.


Relays

  • Alarm Relay s 1 summary alarm for ” analyzer failure”Maximum load:1A / 24 V
  • 2 potential-free contacts for each channel programmable as limit switches for measuring values (high/low thresholds)
  • 1 sample flow alarm for each channel
  • 1 output for indication of the active sample stream for each channel.
  • 1 output for maintenance indication.


For full details please view or download the Dissolved Iron DATE SHEET

  • Dissolved Iron in Potable Water.
  • Dissolved Iron in Wastewater
  • Dissolved Iron in Surface Water
  • Dissolved Iron in Effluents

Analyzer

  • Seres OL Topaz Iron LR (0 – 100 ppb or 0 – 1000 ppb) SOL–55.331.000
  • Seres OL Topaz Iron HR (0 – 2 ppm or 0 – 5 ppm) SOL–55.331.100

Configuration

  • 2-Channel Setup (similar range) (similar range) SOL-83.590.020
  • 4-Channel Setup (similar range) (similar range) SOL-83.590.040
  • 6-Channel Setup (similar range) (similar range) SOL-83.590.060
  • Ethernet Interface (TCP/IP) SOL-81.410.020

Option

  • 1-Year Spare Part Package “Basis” (Analyzer + 1st channel) SOL-84.110.080
  • 1-Year Spare Part Package “Multi-Channel” (add once if multi-channel configuration was selected) SOL-84.110.150
  • Reagent shelf inf SS316L SOL-89.610.010

Reagent specifications

  • pH buffer acetate pH 4.5 RXX237
    Reagent Consumption LR/HR 0.6l/month
  • Phenanthroline 1.10 RXX151
    Reagent Consumption LR 1.6 l/month
    Reagent Consumption HR 2.4 l/month
  • Topaz Iron LR
    • Hydrochloric acid 2N
      + 0.4mg/l iron RXX153MI400
      Reagent Consumption 0.6l/month
  • Topaz Iron HR
    • Hydrochloric acid 2N RXX150
      Reagent Consumption 0.6l/month
  • 2 programmable signal outputs for measured values (freely scalable, linear). Current loop:4 – 20 mA
  • Communication interface RS485 interface (galvanically separated) withModbus/JBUS RTU protocol.
  • Ethernet interface (TCP/IP) optional.

Relays

  • Alarm Relay s 1 summary alarm for ” analyzer failure”Maximum load:1A / 24 V
  • 2 potential-free contacts for each channel programmable as limit switches for measuring values (high/low thresholds)
  • 1 sample flow alarm for each channel
  • 1 output for indication of the active sample stream for each channel.
  • 1 output for maintenance indication.

All Swan Analyzers are CSA Certified at our premises before shipping.

Total Dissolved Iron Anlysis

Complete iron analyzer system for the automatic, continuous analysis of dissolved iron Fe (II+III) in potable water, surface water, wastewater and effluents.

For the continuous, colorimetric online determination of dissolved iron Fe (II+III) per ISO 6332:1988, DIN 38406-E1-1 and NF T90 017.

  • Available in separate measuring ranges:
    • Seres OL Topaz Iron LR:
      • 0 – 100 ppb or
      • 0 – 1000 ppb
    • Seres OL Topaz Iron HR:
      • 0 – 2 ppm or
      • 0 – 5 ppm
  • Complete iron analyzer including measurement and control electronics, measuring unit, flow indicator, reaction chamber and reagent dosing system.
  • Robust, high quality analyzer cabinet painted stainless steel, 316L.
  • Automatic, electrical zero measurement prior to each measurement cycle.
  • Automatic cell cleaning.
  • 3 easily accessible peristaltic pump modules for accurate, automatic dosing of chemical reagents.
  • 2 analog and 7 digital outputs for alarms for process values and diagnostic alarms for each sample stream.
  • RS485 Modbus/JBUS RTU interface.
  • Large back-lit touchscreen color LCD display for the reading of all measured values and status information simultaneously.
  • Easy menu-guided operation in English or French.

The Serez OL Topaz series provides you with a high-quality industrial-grade analyzer specific to your Dissolved Iron analysis requirement.

Fluoride

Colorimetric method; in a medium buffered to pH 4.6, iron III is reduced to iron II. Phenanthroline then forms a complex with iron II of red color which is detected by the iron analyzer.

  • Reaction time 15-18 min.
  • Topaz Iron LR 0-100 or 0-1000 ppb
    • Limit of Detection 8 ppb
    • Repeatability < ± 2 % FS or ± 0.008ppm (whichever is greater)
    • Precision < ± 2 % FS or ± 0.008ppm (whichever is greater)
  • Topaz Iron HR 0-2 or 0-5 ppm
    • Limit of Detection 8 ppb
    • Repeatability < ± 2 % FS
    • Precision < ± 2 % FS


Complete iron analyzer including measurement and control electronics, measuring unit, flow indicator, reaction chamber and reagent dosing system.

Signal outputs

  • 2 programmable signal outputs for measured values (freely scalable, linear).
  • Current loop: 4 – 20 mA
  • Communication interfaceRS485 interface (galvanically separated) with Modbus/JBUS RTU protocol.
  • Ethernet interface (TCP/IP) optional.


Relays

  • Alarm Relay s 1 summary alarm for ” analyzer failure”Maximum load:1A / 24 V
  • 2 potential-free contacts for each channel programmable as limit switches for measuring values (high/low thresholds)
  • 1 sample flow alarm for each channel
  • 1 output for indication of the active sample stream for each channel.
  • 1 output for maintenance indication.


For full details please view or download the Dissolved Iron DATE SHEET

  • Dissolved Iron in Potable Water.
  • Dissolved Iron in Wastewater
  • Dissolved Iron in Surface Water
  • Dissolved Iron in Effluents

Analyzer

  • Seres OL Topaz Iron LR (0 – 100 ppb or 0 – 1000 ppb) SOL–55.331.000
  • Seres OL Topaz Iron HR (0 – 2 ppm or 0 – 5 ppm) SOL–55.331.100

Configuration

  • 2-Channel Setup (similar range) (similar range) SOL-83.590.020
  • 4-Channel Setup (similar range) (similar range) SOL-83.590.040
  • 6-Channel Setup (similar range) (similar range) SOL-83.590.060
  • Ethernet Interface (TCP/IP) SOL-81.410.020

Option

  • 1-Year Spare Part Package “Basis” (Analyzer + 1st channel) SOL-84.110.080
  • 1-Year Spare Part Package “Multi-Channel” (add once if multi-channel configuration was selected) SOL-84.110.150
  • Reagent shelf inf SS316L SOL-89.610.010

Reagent specifications

  • pH buffer acetate pH 4.5 RXX237
    Reagent Consumption LR/HR 0.6l/month
  • Phenanthroline 1.10 RXX151
    Reagent Consumption LR 1.6 l/month
    Reagent Consumption HR 2.4 l/month
  • Topaz Iron LR
    • Hydrochloric acid 2N
      + 0.4mg/l iron RXX153MI400
      Reagent Consumption 0.6l/month
  • Topaz Iron HR
    • Hydrochloric acid 2N RXX150
      Reagent Consumption 0.6l/month
  • 2 programmable signal outputs for measured values (freely scalable, linear). Current loop:4 – 20 mA
  • Communication interface RS485 interface (galvanically separated) withModbus/JBUS RTU protocol.
  • Ethernet interface (TCP/IP) optional.

Relays

  • Alarm Relay s 1 summary alarm for ” analyzer failure”Maximum load:1A / 24 V
  • 2 potential-free contacts for each channel programmable as limit switches for measuring values (high/low thresholds)
  • 1 sample flow alarm for each channel
  • 1 output for indication of the active sample stream for each channel.
  • 1 output for maintenance indication.

All Swan Analyzers are CSA Certified at our premises before shipping.