Seres Topaz Aluminum

For the continuous, colorimetric online determination of aluminum per DIN ISO 10566:1994 within ranges of 0 to 200 ppb.

Aluminum Analyzer

Online Aluminum Analyzer

Complete monitoring system for the automatic, continuous measurement of aluminum in potable water, wastewater and effluents.

Features of the Topaz online aluminum analyzer include:
For the continuous, colorimetric online determination of aluminum per DIN ISO 10566:1994 within ranges of 0 to 200 ppb.

  • Complete system including measurement and
    control electronics, measuring unit, flow indicator,
    reaction chamber and reagent dosing system.
  • Robust, high quality analyzer cabinet painted stainless
    steel, 316L.
  • Automatic, electrical zero measurement prior to each
    measurement cycle.
  • Automatic cell cleaning.
  • 4 easily accessible peristaltic pump modules for
    accurate, automatic dosing of chemical reagents.
  • 2 analog and 7 digital outputs for alarms for process
    values and diagnostic alarms for each sample stream.
  • RS485 Modbus/JBUS RTU interface.
  • Large back-lit touchscreen color LCD display for the
    reading of all measured values and status information
    simultaneously.
  • Easy menu-guided operation in English or French

The Serez OL Topaz series provides you with a high-quality industrial-grade analyzer specific to your aluminum analysis requirement.

Aluminum Analyzer
Colorimetric determination of aluminum ions after chemical reaction in an alkaline environment.:
  • Reaction time <13 min
  • Topaz Aluminum 0-200 ppb
    • Limit of Detection 6 ppb
    • Repeatability ± 2 % FS or ± 3 ppb (whichever is greater)
    • Accuracy ± 3 % FS or ± 3 ppb (whichever is greater)
Complete system including measurement and control electronics, measuring unit, flow indicator, reaction chamber and reagent dosing system.

Signal outputs
  • 2 programmable signal outputs for measured values (freely scalable, linear).
  • Current loop: 4 – 20 mA
  • Communication interfaceRS485 interface (galvanically separated) with Modbus/JBUS RTU protocol.
  • Ethernet interface (TCP/IP) optional.
Relays
  • Alarm Relay s 1 summary alarm for ” analyzer failure”Maximum load:1A / 24 V
  • 2 potential-free contacts for each channel programmable as limit switches for measuring values (high/low thresholds)
  • 1 sample flow alarm for each channel
  • 1 output for indication of the active sample stream for each channel.
  • 1 output for maintenance indication.
For full details please view or download the DATE SHEET
  • Aluminum in Potable Water
  • Aluminum in Wastewater
  • Aluminum in Effluents

Configuration

  • 2-Channel Setup (LR only, similar range) SOL-83.590.020
  • 4-Channel Setup (LR only, similar range) SOL-83.590.040
  • 6-Channel Setup (LR only, similar range) SOL-83.590.060
  • Ethernet Interface (TCP/IP) SOL-81.410.020

Option

  • 1-Year Spare Part Package “Basis” (Analyzer + 1st channel) SOL-84.110.040
  • 1-Year Spare Part Package “Multi-Channel” (add once if multi-channel config. was selected) SOL-84.110.150
  • Reagent Shelf in SS316L SOL-89.610.010

Reagent specifications

  • Chlorhydric acid 2N +400μg/L Aluminum RXX152
    Reagent Consumption 0.6l/month
  • Phenanthroline -1,10 RXX151
    Reagent Consumption 0.8l/month
  • Pyrocatechol violet RXX194
    Reagent Consumption 0.8l/month
  • Buffer pH 6-6.2 Hexamethylenetetramine RXX193
    Reagent Consumption 3.3l/month
  • 2 programmable signal outputs for measured values (freely scalable, linear). Current loop:4 – 20 mA
  • Communication interface RS485 interface (galvanically separated) withModbus/JBUS RTU protocol.
  • Ethernet interface (TCP/IP) optional.

Relays

  • Alarm Relay s 1 summary alarm for ” analyzer failure”Maximum load:1A / 24 V
  • 2 potential-free contacts for each channel programmable as limit switches for measuring values (high/low thresholds)
  • 1 sample flow alarm for each channel
  • 1 output for indication of the active sample stream for each channel.
  • 1 output for maintenance indication.

All Swan Analyzers are CSA Certified before shipping.

Description

Online Aluminum Analyzer

Complete monitoring system for the automatic, continuous measurement of aluminum in potable water, wastewater and effluents.

Features of the Topaz online aluminum analyzer include:
For the continuous, colorimetric online determination of aluminum per DIN ISO 10566:1994 within ranges of 0 to 200 ppb.

  • Complete system including measurement and
    control electronics, measuring unit, flow indicator,
    reaction chamber and reagent dosing system.
  • Robust, high quality analyzer cabinet painted stainless
    steel, 316L.
  • Automatic, electrical zero measurement prior to each
    measurement cycle.
  • Automatic cell cleaning.
  • 4 easily accessible peristaltic pump modules for
    accurate, automatic dosing of chemical reagents.
  • 2 analog and 7 digital outputs for alarms for process
    values and diagnostic alarms for each sample stream.
  • RS485 Modbus/JBUS RTU interface.
  • Large back-lit touchscreen color LCD display for the
    reading of all measured values and status information
    simultaneously.
  • Easy menu-guided operation in English or French

The Serez OL Topaz series provides you with a high-quality industrial-grade analyzer specific to your aluminum analysis requirement.

Aluminum Analyzer
Specifications
Colorimetric determination of aluminum ions after chemical reaction in an alkaline environment.:
  • Reaction time <13 min
  • Topaz Aluminum 0-200 ppb
    • Limit of Detection 6 ppb
    • Repeatability ± 2 % FS or ± 3 ppb (whichever is greater)
    • Accuracy ± 3 % FS or ± 3 ppb (whichever is greater)
Complete system including measurement and control electronics, measuring unit, flow indicator, reaction chamber and reagent dosing system.

Signal outputs
  • 2 programmable signal outputs for measured values (freely scalable, linear).
  • Current loop: 4 – 20 mA
  • Communication interfaceRS485 interface (galvanically separated) with Modbus/JBUS RTU protocol.
  • Ethernet interface (TCP/IP) optional.
Relays
  • Alarm Relay s 1 summary alarm for ” analyzer failure”Maximum load:1A / 24 V
  • 2 potential-free contacts for each channel programmable as limit switches for measuring values (high/low thresholds)
  • 1 sample flow alarm for each channel
  • 1 output for indication of the active sample stream for each channel.
  • 1 output for maintenance indication.
For full details please view or download the DATE SHEET
Application
  • Aluminum in Potable Water
  • Aluminum in Wastewater
  • Aluminum in Effluents
Options

Configuration

  • 2-Channel Setup (LR only, similar range) SOL-83.590.020
  • 4-Channel Setup (LR only, similar range) SOL-83.590.040
  • 6-Channel Setup (LR only, similar range) SOL-83.590.060
  • Ethernet Interface (TCP/IP) SOL-81.410.020

Option

  • 1-Year Spare Part Package “Basis” (Analyzer + 1st channel) SOL-84.110.040
  • 1-Year Spare Part Package “Multi-Channel” (add once if multi-channel config. was selected) SOL-84.110.150
  • Reagent Shelf in SS316L SOL-89.610.010

Reagent specifications

  • Chlorhydric acid 2N +400μg/L Aluminum RXX152
    Reagent Consumption 0.6l/month
  • Phenanthroline -1,10 RXX151
    Reagent Consumption 0.8l/month
  • Pyrocatechol violet RXX194
    Reagent Consumption 0.8l/month
  • Buffer pH 6-6.2 Hexamethylenetetramine RXX193
    Reagent Consumption 3.3l/month
Downloads
F.A.Q
  • 2 programmable signal outputs for measured values (freely scalable, linear). Current loop:4 – 20 mA
  • Communication interface RS485 interface (galvanically separated) withModbus/JBUS RTU protocol.
  • Ethernet interface (TCP/IP) optional.

Relays

  • Alarm Relay s 1 summary alarm for ” analyzer failure”Maximum load:1A / 24 V
  • 2 potential-free contacts for each channel programmable as limit switches for measuring values (high/low thresholds)
  • 1 sample flow alarm for each channel
  • 1 output for indication of the active sample stream for each channel.
  • 1 output for maintenance indication.

All Swan Analyzers are CSA Certified before shipping.