High precision monitor with three electrode Oxytrace G sensor for the continuous measurement of trace levels of dissolved oxygen in high purity water.
Continuous amperometric measurement of trace levels of dissolved Oxygen in high purity water.
Swansensor Oxytrace G with three electrode set-up (cathode [gold], anode [silver] and guard [silver]) with integrated NT5k temperature sensor.
Complete system mounted on stainless steel mounting panel.
Transmitter in aluminum housing (IP 66) with 2 signal outputs for measuring value or for continuous control, 2 relays as limit switch or controller or timer, 1 alarm relay as summary alarm contact, 1 input for potential-free contact.
Flow cell QV-Flow PMMA OTG made of acrylic glass with needle valve and digital sample flow meter on mounting angle made of stainless steel.
| Monitor AMI Oxytrace – Trace Levels of Dissolved Oxygen | |
| Measuring range | Measuring range with Swansensor OXYTRACE G (at 25 °C): – 0 – 20.0 ppm O2 or – 0 – 200% saturation |
| Transmitter Specifications | |
| Power Supply |
|
| Power consumption | max. 35 VA |
| Sample conditions | |
| Water consumption | 8 to 25 l/h |
| Water inlet pressure | 0.2 to 1 bar |
| Sample temperature | up to 45 °C |
| Panel | |
| Dimension | 280 x 850 x 150 mm |
| Material | stainless steel |
| Weight | 8 kg |
AMI Oxytrace Data Sheet: Trace Levels of Dissolved Oxygen .pdf
All Swan Analyzers are CSA Certified before shipping.
High precision monitor with three electrode Oxytrace G sensor for the continuous measurement of trace levels of dissolved oxygen in high purity water.
Continuous amperometric measurement of trace levels of dissolved Oxygen in high purity water.
Swansensor Oxytrace G with three electrode set-up (cathode [gold], anode [silver] and guard [silver]) with integrated NT5k temperature sensor.
Complete system mounted on stainless steel mounting panel.
Transmitter in aluminum housing (IP 66) with 2 signal outputs for measuring value or for continuous control, 2 relays as limit switch or controller or timer, 1 alarm relay as summary alarm contact, 1 input for potential-free contact.
Flow cell QV-Flow PMMA OTG made of acrylic glass with needle valve and digital sample flow meter on mounting angle made of stainless steel.
| Monitor AMI Oxytrace – Trace Levels of Dissolved Oxygen | |
| Measuring range | Measuring range with Swansensor OXYTRACE G (at 25 °C): – 0 – 20.0 ppm O2 or – 0 – 200% saturation |
| Transmitter Specifications | |
| Power Supply |
|
| Power consumption | max. 35 VA |
| Sample conditions | |
| Water consumption | 8 to 25 l/h |
| Water inlet pressure | 0.2 to 1 bar |
| Sample temperature | up to 45 °C |
| Panel | |
| Dimension | 280 x 850 x 150 mm |
| Material | stainless steel |
| Weight | 8 kg |
AMI Oxytrace Data Sheet: Trace Levels of Dissolved Oxygen .pdf
All Swan Analyzers are CSA Certified before shipping.